This article is also available in 简体中文.
Short answer: the two numbers are not the same kind of number and cannot be equated. "400X" is an angular magnification referred to the human eye — objective × eyepiece. A zoom lens datasheet states object-side magnification β, the sensor image ÷ the real object. No arithmetic maps "400X" onto a lens label; only the field of view does. Fix the FOV and the sensor and β follows; the apparent on-screen figure then follows from the display.
This page covers the conversion: reading the 400X label, the display-magnification chain, a conversion table against real POMEAS models, and the four sources of conversion error. Selection is handled by From Microscope to Machine Vision: How to Select a Camera, Zoom Lens and Magnification.
Part of the POMEAS Machine Vision Optics Comparisons series — see the Machine Vision Optics Comparisons index for all 15 articles in this set, grouped by the option being compared.
A compound microscope enlarges in two stages: objective × eyepiece. A bench microscope reads 40X × 10X = 400X, and two properties of that label decide the rest.
It is referenced to the unaided eye. 400X is an angular comparison: the image through the eyepiece subtends about 400 times the angle the object subtends at a 250 mm reference distance — a viewing experience, not a sensor.
The eyepiece contributes 10X of it and no resolving power. Resolution is set by the objective, chiefly its numerical aperture. Put a camera at the same intermediate image plane and nothing optical is lost — the reference moves from an eye to a sensor plane.
And 400X says nothing about field of view. In microscope practice, object FOV ≈ eyepiece field number ÷ objective magnification. With field number 18: 18 ÷ 40 = 0.45 mm (calculated from the formula, conditional on that field number). A field number 26 eyepiece gives ≈0.65 mm at the same 400X — which is why the conversion has to start from FOV, not from "X".
Almost every failed conversion comes from mixing these up — different reference planes, different owners, only one usable for lens selection.
| Quantity | Definition / reference plane | On a lens datasheet? | Usable for selection? |
|---|---|---|---|
| Objective magnification (40X in the example) | Objective alone; resolution set by its NA | On the objective barrel only | No — sets resolution, not FOV |
| Total optical (visual) magnification (400X) | Objective × eyepiece; angular, to the unaided eye | On the microscope label only | No — no sensor in this chain |
| Object-side optical magnification β | Sensor image ÷ object; in practice sensor width ÷ FOV | Yes — the only one a datasheet guarantees | Yes |
| Electronic / display magnification | β × (monitor width ÷ sensor width) × software zoom | No — set by your monitor and software | Operator acceptance only |
The two figures sit on different reference planes — which is why "which zoom lens equals 400X?" has no answer. Only the FOV is shared; for why a lens "X" and a system "X" differ, see Zoom vs Magnification: What Is the Difference?
The display step is where the quoted number usually comes from, and the least reliable one — it changes when someone swaps a monitor:
β = sensor active width ÷ required horizontal FOV
M_display ≈ β × (monitor display width ÷ sensor active width) × software zoom
2/3″ camera (8.8 mm active width, the format width used in POMEAS FOV tables), 16 mm max horizontal FOV, 24″ 16:9 monitor (≈531 mm display width), no software zoom.
That last figure is the honest version of "what equals 400X?" — a display-matching number, never a resolution guarantee.
| Monitor diagonal | M_display at β = 0.55X | β required for 400X on screen |
|---|---|---|
| 21.5″ (≈476 mm wide) | ≈29.7X | ≈7.4X |
| 24″ (≈531 mm wide) | ≈33.2X | ≈6.6X |
| 27″ (≈598 mm wide) | ≈37.4X | ≈5.9X |
| 32″ (≈708 mm wide) | ≈44.3X | ≈5.0X |
The 12.5X class (PMS-LZL-12101, 0.58X–7.5X) covers all four monitors; the 6.5X class (PMS-LZ-650104, 0.7X–4.5X) covers none. And at a fixed 0.55X, going from a 21.5″ to a 32″ monitor raises apparent magnification by about 50% with no optical change — which is why "it was 400X on the old screen" is not an optical specification.
Two routes behave differently against a "400X" requirement. Fixed-magnification telecentric lenses hold β, distortion and WD across the whole field — the closest metrology-grade substitute for an objective. Continuous zoom lenses cover a range, for inspections needing both a wide find-it and a fine judge-it field.
| Model | Type | Optical magnification | Stated resolution | WD | Horizontal FOV on 2/3″ | Distortion |
|---|---|---|---|---|---|---|
| PMS-LTCM15-110C | Coaxial telecentric, fixed | 1.5X | 5.59 µm | 110 mm | ≈5.9 mm (calculated) | 0.01% |
| PMS-LTCM12-110C | Coaxial telecentric, fixed | 1.2X | 6.71 µm | 110 mm | ≈7.3 mm (calculated; official FOV = 7.3 × 5.5 mm) | 0.01% |
| PMS-LTC2-110C | Telecentric, fixed | 2.0X | 8 µm | 110 mm | ≈4.4 mm (calculated) | 0.01% |
| PMS-LZ-650104 | Motorized zoom, 6.5:1 | 0.7X–4.5X | — | 82 ± 2 mm | ≈12.6 mm at 0.7X → ≈2.0 mm at 4.5X (calculated) | — |
| PMS-LZL-12101 | Manual coaxial zoom, 12.5X, telecentric design | 0.58X–7.5X | minimum 3.05 µm | 77.4 ± 2 mm | ≈15.2 mm at 0.58X → ≈1.2 mm at 7.5X (calculated) | minimum TV distortion 0.001% |
The core point of this page: no model in this table "is" a 400X microscope. The 0.55X result sits below the wide end of every model listed; the 6.6X display-matching result sits inside the PMS-LZL-12101 range and outside the PMS-LZ-650104 range. Models are matched to the FOV and resolution requirement, never to the microscope's number — put the FOV, sensor format and resolution on the specification sheet, not "equivalent to 400X".
The formulas are in the Zoom Lens Magnification & FOV Calculator; the fixed-magnification route is compared model by model in the telecentric lens selection guide.
A correct calculation can still be 5–20% off once hardware and software are in the loop; four sources account for most of it.
| # | Source | Mechanism and effect | Self-check |
|---|---|---|---|
| 1 | Magnification at the field radius | Conventional optics vary magnification from centre to edge, so the same feature measures differently at centre and edge; telecentric designs keep the chief ray parallel to the axis | Image a calibration target and derive mm/pixel at centre and four corners. Reference: 0.01% distortion on PMS-LTCM15-110C / PMS-LTCM12-110C; 0.001% TV distortion on PMS-LZL-12101 |
| 2 | Distortion | Radial and tangential error curves straight lines, so nominal β is right at the centre and wrong where the defect sits | Image a straight-edged gauge; if the edge bows, calibrate per zone rather than using one global mm/pixel figure |
| 3 | Display scaling and aspect ratio | The monitor stretches the sensor image to fit and software zoom adds an arbitrary factor, so M_display changes with no optical change | Lock the display mode and software zoom at 1.0× in the recipe; re-derive M_display after any monitor change |
| 4 | Non-telecentric optics, WD variation | β depends on object distance, so part height and focus move it — magnification is no longer constant | Move a gauge through the expected height range and record the measured size at each level |
Reference points: the PMS-LZL-12101 12.5X coaxial continuous zoom lens (0.58X–7.5X) covers the 6.6X display-matching point; the PMS-LZ-650104 (0.7X–4.5X) covers wider-field work. Where β must be fixed and distortion must not vary across the field, the fixed coaxial telecentric models are the metrology route: PMS-LTCM15-110C (1.5X), PMS-LTCM12-110C (1.2X) and PMS-LTC2-110C (2.0X).
The same caution applies when a microscope’s 2000× is quoted against a camera system: what 2000× actually means for semiconductor observation separates object-side from display magnification.
Send six numbers: measured FOV (worth more than the magnification label), camera model or sensor format, max and min FOV, smallest feature, working distance, and manual or automated inspection. POMEAS can evaluate the required object-side magnification, the zoom range, the display-magnification figure and the resolution checks that decide whether a zoom lens alone is sufficient — Contact POMEAS engineering.
There is no equivalent magnification number, because the two figures use different reference planes. 400X is an angular magnification referred to the human eye (objective × eyepiece); a lens datasheet states object-side magnification β, the ratio of the sensor image to the real object. The conversion has to go through the field of view. If the requirement is a 16 mm horizontal field on a 2/3″ sensor (8.8 mm active width), β = 8.8 ÷ 16 ≈ 0.55X, calculated from the formula — and that value must be re-derived for any other sensor format.
Optical total magnification is objective × eyepiece — an angular magnification referred to the unaided eye, and the only stage that produces new optical detail. Display magnification is β × (monitor display width ÷ sensor active width) × software zoom: it enlarges an image the sensor has already captured and adds no information. A 0.55X lens on a 2/3″ sensor with a 24″ monitor gives about 33X on screen (calculated from the formula); the same optical chain on a 21.5″ monitor gives about 29.7X. The optics did not change — only the display term did.
Two steps. First, β = sensor active width ÷ required FOV. Second, M_display ≈ β × (monitor display width ÷ sensor active width) × software zoom. For a 2/3″ sensor (8.8 mm) at β = 0.55X and a 24″ 16:9 monitor (≈531 mm wide): M_display ≈ 0.55 × 531 ÷ 8.8 ≈ 33X, calculated from the formula. Note that this is a linear size ratio on screen, not the microscope's angular convention; comparing it to a microscope figure rigorously would also require the viewing distance and the display pixel pitch.
Rearranging the display formula: β = 400 × sensor active width ÷ monitor display width. On a 2/3″ sensor (8.8 mm) with a 24″ monitor (≈531 mm) that is ≈6.6X (calculated from the formula); a 21.5″ monitor pushes it to ≈7.4X and a 32″ monitor lowers it to ≈5.0X. The 12.5X class (PMS-LZL-12101, 0.58X–7.5X) covers all three monitor sizes; the 6.5X class (PMS-LZ-650104, 0.7X–4.5X) covers none. Remember this is a display-matching figure, not a resolution guarantee.
Four common causes, in order of frequency: magnification varying out toward the field radius (conventional optics versus telecentric design), distortion bending straight edges, display scaling or software zoom not locked, and β drifting with working distance in a non-telecentric lens. Self-check by imaging a calibration target and deriving mm/pixel at the centre and the four corners. Reference figures for well-corrected models: PMS-LTCM15-110C distortion 0.01%, PMS-LTCM12-110C distortion 0.01% with object-side telecentricity below 0.02°, and PMS-LZL-12101 minimum TV distortion 0.001%.
It depends on whether the requirement is one field or a range. If the inspection works at a single magnification and the measurement must stay stable across the whole field, a fixed-magnification coaxial telecentric lens is the closer substitute for an objective — for example PMS-LTCM15-110C (1.5X, 5.59 µm, WD 110 mm), PMS-LTCM12-110C (1.2X, 6.71 µm, WD 110 mm) or PMS-LTC2-110C (2.0X, 8 µm, WD 110 mm), all 2/3″ with 0.01% distortion. If the task needs both a wide find-it field and a fine judge-it field, a continuous zoom covers the range — PMS-LZL-12101 (0.58X–7.5X, minimum resolution 3.05 µm, WD 77.4 ± 2 mm) or PMS-LZ-650104 (0.7X–4.5X, WD 82 ± 2 mm).
The hardware referred to above is grouped into the following categories on the product side, with the specifications and model list for each range.
The comparison above sets out why screen magnification and optical magnification are different quantities. This section adds the figures that the POMEAS microscope units publish, so the difference can be quantified rather than described. Across those units magnification spans 1 to 2 X (n=2) and field of view spans 10 to 360 mm (n=3) at the corresponding configurations. The field figures are the ones that matter for deciding whether a part fits in one view.
The figure that matters for measurement is the total magnification, which includes the objective, the tube lens and any adapter factor. A screen figure that omits one of these overstates what the sensor actually receives. Because the total is a product, a small error in the adapter factor scales the whole result. This is why the adapter should be part of the specified configuration rather than an accessory chosen later.
Higher magnification reduces depth of field, so the fixture flatness requirement tightens as magnification rises. A station that works at low magnification on a stamped part may need a completely different presentation at high magnification. The measured working distance range for these units is 22 to 55 mm (n=2), which is modest because the units are complete assemblies with an illumination path inside the body.
For visual inspection and for operator training, the screen figure is the honest one, because it describes what the operator sees. For measurement and for defect detection algorithms, only the optical figure at the sensor matters. Mixing the two in a specification produces a requirement that no configuration can satisfy, because it asks for screen-level detail at sensor-level sampling. Stating which figure the requirement refers to, at the start of the specification, prevents that confusion.
Simply enter your email to receive the latest news and insights from Pomeas. Stay connected with Pomeas and be the first to discover new innovations in optical excellence.