"Microscope" covers four different optical paths in an industrial setting, and they are not interchangeable. A conventional microscope with eyepieces and a turret of objectives, a motorised zoom body driving a camera, a laser autofocus system that holds focus on a moving surface, and a white-light interferometer measuring nanometre-scale surface structure are four answers to four different questions. Choosing between them starts with naming what the inspection has to output.
Path 1 — Conventional industrial microscope. Objectives on a turret, infinity-corrected optics, coaxial illumination, observed through eyepieces or a camera port. This is the configuration used for visual inspection, documentation, and manual judgement. Its strength is the operator's ability to explore the sample freely at high magnification; its limit is that the output is an image, not a number, unless paired with measuring software.
POMEAS builds infinity-corrected microscope units with coaxial illumination in three formats: MPT-AF (large-target coaxial illumination infinity system), MPT-BF (full-frame coaxial illumination infinity system), and MPT-N (C-mount infinity coaxial system). The three differ in sensor coverage and interface, which is what decides which one pairs with a given camera.
Path 2 — Motorised zoom microscope. A zoom body provides continuous magnification change under motor control, and autofocus keeps the image sharp as magnification or sample height changes. This is the path for automated inspection, because the magnification becomes a parameter in a recipe rather than a mechanical adjustment.
POMEAS PMS-MF6020 is a motorised zoom autofocus microscope, and MICRO IMAGE3 is a 4K automatic digital microscope system specified for 3D imaging and precision measurement.
Path 3 — Laser autofocus microscopy. Where the surface height varies across the sample or the part moves, maintaining focus by image contrast is slow and fails on low-contrast surfaces. A coaxial laser autofocus channel measures distance optically and drives the focus mechanism directly.
POMEAS PMS-FocusX is specified with motorised objective switching and coaxial laser focus. The combination matters: switching objectives changes the working distance, and a coaxial laser channel re-establishes focus without a contrast search.
Path 4 — White-light interferometry. When the output is surface roughness or step height at the nanometre scale, imaging optics are the wrong tool. WLI 1000 is specified as a nanometre-scale 3D white-light interference profilometer for surface roughness and step height measurement.
A recurring problem in industrial microscopy is a sample that is both three-dimensional and needs to be sharp everywhere at once — a metal surface with scratches and burrs, a fracture face, a textured coating. A single exposure at high magnification has a shallow depth of field and cannot hold the whole surface in focus.
PMS-56665 is specified as an extended-depth-of-field (超景深) industrial digital microscope for metal surface scratch and burr inspection using focus stacking (景深合成). Focus stacking acquires a series of images at different focal planes and combines the in-focus regions, producing a single image sharp across a depth no single exposure could hold.
This is the answer when the requirement is "see the whole surface sharply," and it is distinct from the interferometer's requirement, which is "measure the height of the surface numerically."
| Requirement | Optical path | POMEAS reference |
|---|---|---|
| Visual inspection, documentation, operator judgement | Conventional microscope, infinity optics | MPT-AF / MPT-BF / MPT-N |
| Automated inspection, magnification in a recipe | Motorised zoom autofocus | PMS-MF6020 |
| Automated 3D imaging and precision measurement at 4K | Automatic digital microscope system | MICRO IMAGE3 |
| Variable sample height, low-contrast surfaces, objective switching | Laser autofocus | PMS-FocusX |
| Whole 3D surface sharp in one image | Extended depth of field, focus stacking | PMS-56665 |
| Surface roughness, step height at nanometre scale | White-light interferometry | WLI 1000 |
Sensor and interface. Infinity microscope units are specified for a sensor format and a camera interface. MPT-N is the C-mount infinity coaxial variant; MPT-AF and MPT-BF are specified for larger targets. Pairing a unit built for a smaller format with a larger sensor produces the same vignetting problem seen in FA lens selection.
Magnification total. On a modular microscope the total magnification is the product of the zoom body, the tube lens, the objective and the adapter factor:
Total magnification = zoom body × tube lens × objective × 0.55
Changing any one term changes the total even when the others are untouched, which is why a microscope's advertised magnification range should always be read together with the objective and tube in use.
Whether the output is an image or a number. If the station must report a dimension, the optical path needs calibration and a distortion figure, not only magnification. If it must report roughness or step height numerically, imaging optics do not apply and an interferometer does.
Whether the surface stays in focus. Sample height variation is handled by laser autofocus (measurement-driven) or by focus stacking (acquisition-driven). The first is faster for a single plane that moves; the second is needed when the surface itself is three-dimensional.
Imaging microscopes, however configured, produce images. Roughness figures, step heights in the nanometre range, and true 3D surface topography are measured by interferometry, not inferred from a sharp image. Conversely, an interferometer is not the tool for looking at a scratch and judging whether it is a defect.
Neither path removes the need for illumination discipline. Coaxial illumination helps on specular surfaces; the infinity units above are specified with coaxial illumination for this reason. On a rough metal surface, the illumination angle decides whether a scratch appears as a dark line or disappears entirely.
For automated inspection, yes in function: magnification becomes a recipe parameter. For exploratory manual inspection, the conventional path remains the practical choice.
Holding focus when the sample height varies or the surface has low contrast. It measures distance optically rather than searching for image contrast.
A single image sharp across a depth greater than one exposure's depth of field, by combining in-focus regions from a focal series.
When the output is a number for roughness or step height at nanometre scale. Imaging optics do not measure that.
Zoom body × tube lens × objective × 0.55 (adapter factor). Changing any term changes the total.