Quick answer: A missing 1 µm defect on a SiC wafer is usually a resolution problem, not a pixel problem. At its 7.5× end, the POMEAS PMS-LZL-12104-D1 12.5X motorized zoom reaches a sampling rate of 0.46 µm/pixel, yet its datasheet optical resolution at that same magnification is 3.05 µm (N.A. 0.11). Sampling is not resolution. To actually resolve 1 µm you need an APO microscope objective — the PMS-MPO10 is specified at 1 µm — and you must then deal with depth of field collapsing from 50 µm to 3.5 µm, which is what laser autofocus (PMS-FocusX, ≤ 0.4 s cycle) exists to solve.
Silicon carbide device yield is largely a function of substrate and epi-layer defect control. The industry separates SiC defects into two families:
No single optical modality captures both families. That is why production SiC final-inspection tools run brightfield, darkfield and photoluminescence (PL) channels in parallel.
Brightfield uses coaxial (episcopic) epi-illumination: light goes down through the objective and reflects straight back. A polished, flat area returns strong specular reflection and images bright; a scratch, particle or stain breaks that specular return and appears dark. For polished SiC, coaxial illumination is not a nicety — it is the standard way to suppress specular glare off a mirror-like surface.
Darkfield illuminates at a very low grazing angle. Flat surface reflects light away from the objective and the background goes black; only a defect with topography scatters light back into the lens and appears as a bright point on black. Sensitivity to sub-micron particle scattering is materially higher than brightfield, at the cost of a dedicated grazing illuminator, weak signal, and high sensitivity to vibration and cleanliness.
PL excites the SiC with near-UV light. Defective regions emit at different wavelengths and intensities than sound material, which makes dislocations, stacking faults and micropipes visible — defects that neither the eye nor brightfield can see. The method is standardised in China as GB/T 43493.3-2023 (Non-destructive detection and identification criteria for defects in silicon carbide homoepitaxial wafers for power devices — Part 3: Photoluminescence detection method).
| Modality | What it detects | Optical / illumination requirement | POMEAS supply position |
|---|---|---|---|
| Brightfield | Particles, scratches, contamination, edge chipping, pits | Coaxial epi-illumination | ✅ In scope (coaxial microscope units, coaxial zoom lens) |
| Darkfield | Sub-micron particles, shallow scratches, micro-chipping | Dedicated low-angle grazing illuminator | ⚠️ Partial — standard LED lighting only, no wafer-specific darkfield module; engineering confirmation required |
| PL | BPD, TSD, TED, stacking faults, micropipes | Near-UV excitation + emission filter + dichroic | ❌ Out of POMEAS scope |
Capability boundary, stated plainly. POMEAS supplies optical imaging components. The UV excitation source, dichroic mirror and emission filters that a PL channel requires are not POMEAS products, and objective transmission at a 275 nm excitation band must be confirmed separately. POMEAS does not supply a complete SiC wafer inspection tool, nor does it supply defect classification algorithms.
There is exactly one formula:
object-space sampling (µm/pixel) = camera pixel pitch (µm) ÷ optical magnification
Using a 2/3″ 5 MP camera (2448 × 2048, 3.45 µm pixels) on the POMEAS PMS-LZL-12104-D1 12.5X motorized zoom (0.58×–7.5×):
| Magnification | Object-space sampling | Typical use |
|---|---|---|
| 0.58× | 5.95 µm/pixel | Largest field — whole-die / wafer region locating |
| 1.0× | 3.45 µm/pixel | — |
| 3.0× | 1.15 µm/pixel | — |
| 6.0× | 0.575 µm/pixel | — |
| 7.5× | 0.46 µm/pixel | Top end of the zoom range |
Inverting it: reaching 0.5 µm/pixel requires 3.45 ÷ 0.5 ≈ 6.9×. That sits inside the 0.58×–7.5× range, so the claim "a zoom lens can reach 0.5 µm/pixel" is true at the sampling level.
The problem is in the next section.
This is the load-bearing part of the article.
The PMS-LZL-12104-D1 datasheet lists optical performance per magnification. Placed side by side, the contradiction is obvious:
| Magnification | N.A. | Datasheet resolution | Depth of field | Sampling (3.45 µm pixels) |
|---|---|---|---|---|
| 1.0× | 0.043 | 7.8 µm | 0.95 mm | 3.45 µm/pixel |
| 3.0× | 0.09 | 3.73 µm | 0.15 mm | 1.15 µm/pixel |
| 7.5× | 0.11 | 3.05 µm | 0.05 mm (50 µm) | 0.46 µm/pixel |
At the 7.5× stop:
A 1 µm defect is already smeared into a ~3 µm blur spot by diffraction before it ever reaches the sensor, so its contrast collapses and it cannot be detected reliably. Adding pixels does not recover information the optics discarded.
Reliable detection of a 1 µm defect requires two conditions at once:
| Configuration | N.A. | Datasheet resolution | Pixels across 1 µm | Detects 1 µm? |
|---|---|---|---|---|
| 12.5X zoom @ 7.5× | 0.11 | 3.05 µm | ≈ 2.2 px | ❌ No — resolution-limited |
| PMS-MPO10 (10×) | 0.28 | 1 µm | ≈ 2.9 px | ✅ Marginal |
| PMS-MPO20 (20×) | 0.42 | 0.7 µm | ≈ 5.8 px | ✅ Yes |
| PMS-MPO50 (50×) | 0.55 | 0.5 µm | ≈ 14.5 px | ✅ Comfortably |
| Dimension | Motorized zoom lens | APO objective + laser autofocus |
|---|---|---|
| Representative build | PMS-LZL-12104-D1 (12.5X) / PMS-LZH-14105K (14X) | PMS-MPO series + PMS-FocusX |
| Magnification | Continuously variable, 0.58×–7.5× | Fixed steps, motorized turret (2×/5×/10×/20×/50×) |
| Optical resolution | 3.05 µm at 7.5× | 1 µm (MPO10) / 0.7 µm (MPO20) / 0.5 µm (MPO50) |
| Depth of field | 50 µm at 7.5× | 3.5 / 1.6 / 0.9 µm |
| Working distance | 77.4 ± 2 mm (generous) | 34 / 20 / 13 mm (tightens as magnification rises) |
| Best at | Large-field locating → progressive zoom-in review; search and review | Quantitative sub-micron defect detection |
| Weak at | Stable detection of 1 µm-class defects | Fast wide-area search (field only 1.78–0.36 mm) |
| Automation | Motorized zoom, PLC-integrable | RS232 control, ≤ 0.4 s autofocus cycle |
The division of labour in one line: the zoom lens finds and reviews; the objective plus laser autofocus decides whether the 1 µm call is pass or fail. They are stages in one chain, not substitutes.
Moving from a zoom to a high-magnification objective shows up almost entirely as depth of field:
12.5X zoom @ 7.5× DOF 50 µm
PMS-MPO10 DOF 3.5 µm
PMS-MPO20 DOF 1.6 µm
PMS-MPO50 DOF 0.9 µm
Sub-micron depth of field means wafer bow, stage travel error and ordinary floor vibration are each enough to throw the image out of focus. At that point focusing capability, not resolution, decides whether the inspection runs at all.
POMEAS PMS-FocusX laser autofocus microscopy system:
| Parameter | Value |
|---|---|
| Laser focus frequency | 5000 Hz (max) |
| Focus accuracy | 1/4 of depth of field (≈ 0.875 µm at 10×, ≈ 0.4 µm at 20×) |
| Autofocus cycle time | ≤ 0.4 s |
| Z-axis travel | ±15 mm |
| Objective turret | 4-hole / 5-hole motorized, or linear-motor inline |
| Control interface | RS232 |
A second, practical benefit: real-time surface tracking during scanning. This matters for SiC samples that have been defect-selective etched and therefore present real topography — an application where laser focus is generally preferred over image-contrast autofocus.
Supplied (catalogue items, published specifications):
| Category | Model | Key specification |
|---|---|---|
| 12.5X motorized zoom lens | PMS-LZL-12104-D1 | 0.58×–7.5×, WD 77.4 ± 2 mm, 2/3″, C-mount |
| 14X motorized coaxial zoom lens | PMS-LZH-14105K | 0.55×–7.6×, WD 77.8 ± 2 mm, coaxial illumination |
| 10× plan apochromatic objective | PMS-MPO10 | N.A. 0.28, WD 34 mm, 1 µm resolution, 3.5 µm DOF |
| 20× plan apochromatic objective | PMS-MPO20 | N.A. 0.42, WD 20 mm, 0.7 µm resolution, 1.6 µm DOF |
| 50× plan apochromatic objective | PMS-MPO50 | N.A. 0.55, WD 13 mm, 0.5 µm resolution, 0.9 µm DOF |
| Laser autofocus microscopy system | PMS-FocusX | ≤ 0.4 s cycle, 5000 Hz, RS232 |
| Coaxial microscope units | MPT-BF / MPT-AF | 200 mm focal length, F-mount, coaxial illumination |
Not supplied (stated explicitly):
Stated honestly up front: POMEAS has no published SiC wafer inspection case study. The available semiconductor-sector evidence is this project.
For a semiconductor grinding equipment manufacturer, POMEAS configured a system that had to satisfy three simultaneous demands — large-field locating, high-magnification local review, and observation inside a narrow bore:
| System element | Configuration | Role |
|---|---|---|
| Camera | 5 MP, 2/3″, colour CMOS, USB3.0 | Image capture |
| Motorized zoom lens | PMS-LZH-14105K 14X | Large-field locating + high-magnification local review |
| TV tube | 1× TV tube 63010C | Camera-to-zoom interface |
| Motion control | EZC-93104-K | Motorized zoom, integrated into the machine |
| Folded optical path | Prism relay | Viewing inner bores and side walls outside direct line of sight |
| Ring light | GH7430-W(C), Ø40 inner diameter | Surface and edge illumination |
| Light controller | GE24-4PRT 4-channel dimmer | Per-state intensity control |
Key mechanical datums: lens WD 77.8 ± 2 mm, YWD 57 ± 2 mm, XWD 20 ± 2 mm, ring-light standoff approximately 10 ± 5 mm. The inner-bore station carried a ±1 mm height variation, handled by a three-tier strategy: cover it with depth of field where possible, store a separate focus position per station, and add motorized focus / Z-axis / autofocus where neither suffices.
What this case proves: multi-magnification observation, folded optical paths for confined spaces, and height-variation compensation on semiconductor equipment — POMEAS has delivered these. What it does not prove: 1 µm defect detection on SiC wafers. That is a different claim, and without test data it is not made.
None of these can be settled from a datasheet:
Not reliably. At 7.5× the 12.5X zoom samples at 0.46 µm/pixel, but its datasheet optical resolution is 3.05 µm (N.A. 0.11), which is coarser than 1 µm. Sampling is not resolution. For 1 µm detection, move to the PMS-MPO10 (1 µm specified) or higher.
Search and progressive review. The 0.58× end gives a 15.17 × 11.38 mm field for locating; zooming to 7.5× gives 1.17 × 0.88 mm for local review. Production lines typically use the zoom to find and review, and a high-magnification objective to make the final call.
No. The UV excitation source, dichroic mirror and emission filters are not POMEAS products, and POMEAS does not build complete tools. POMEAS supplies the imaging-side optics. For the PL method itself, see GB/T 43493.3-2023.
There is currently no wafer-specific darkfield illumination module in the catalogue — only standard LED bar, spot and point lights. Low-angle grazing darkfield requires a dedicated structure and is an engineering-confirmation item, not a standard-product commitment.
Because depth of field drops to 1.6 µm. Wafer bow, stage error and ambient vibration all cause defocus. PMS-FocusX focuses in ≤ 0.4 s at 1/4 of depth of field (≈ 0.4 µm at 20×) and tracks the surface during scanning.
The 12.5X (PMS-LZL-12104-D1) is 0.58×–7.5× at WD 77.4 ± 2 mm; the 14X (PMS-LZH-14105K) is 0.55×–7.6× at WD 77.8 ± 2 mm and includes coaxial illumination. Choose the 14X when you need to suppress glare on polished surfaces or need coaxial epi-illumination in a confined envelope; choose the 12.5X when you want the lowest distortion and a fully published per-magnification performance table.
Not across series. The zoom lens datasheet uses roughly a 0.61 λ / N.A. criterion; the PMS-MPO objective series uses roughly 0.5 λ / N.A. Cite the stated datasheet value and compare within a series.
Send us the defect family, the smallest defect size you must detect, wafer size, working distance and takt time, and the POMEAS engineering team will respond against the seven-point list above. Wherever 1 µm detection or a PL channel is involved, we will say plainly that sample testing is required rather than promise a number.
Further reading:
All product figures are taken from POMEAS datasheets and pomeas-vision.com product pages; derived values are shown with their formula; SiC defect taxonomy follows published industry method and GB/T 43493.3-2023. Last updated 2026-11-03.