An optical comparator does one thing in a particular way: it projects the silhouette of a part onto a screen and lets a person decide where the edge is, against a chart or overlay, at a stated magnification, under transmitted light. When it has to be second-sourced or replaced, the substitution is not a matter of finding another machine that magnifies. It is a matter of deciding who now decides where the edge is, because that decision is where the error lives.
In the POMEAS global company intelligence set of 2,602 companies, 7 have publicly documented Mitutoyo measuring instruments in use on their own lines. The question is the same on any comparator replacement: the profile is trusted because a human has been reading it for years, and a camera reading the same edge will not agree with a human on every feature, every time. Six parameters decide whether the substitution holds.
A comparator's magnification is a property of its projection lens, and the method assumes the operator knows and trusts it. On a camera-based system there are two numbers and only one is optical. MICRO IMAGE3 publishes both: objective magnification 1X–150X and digital display magnification 1X–8000X. Only the first is a property of the optics; the second is how large the image is drawn.
A zoom system adds a complication. MetX uses a PMS continuous-zoom measuring lens, so magnification is a position, not a constant: the system must be calibrated at the setting actually used, and a recipe that zooms between features inherits the calibration of each position.
The comparator's silhouette exists because light comes from behind the part. On a vision system the same silhouette has to be produced deliberately, and edge quality depends on how collimated that light is: diffuse backlight wraps around an edge and softens it, while a collimated, telecentric-style source keeps the boundary sharp and reduces the sensitivity of the edge position to where the part sits in depth.
POMEAS separates the two channels. MetX carries a high-brightness parallel telecentric cold source for contour work and a five-ring, eight-zone high-brightness white cold source for surface work. IMAGE 3 is configured with surface light, coaxial light and contour backlight, and its documentation states the design reduces perspective error and image distortion so the contour stays stable when a part has height differences. Light a part from the front when the old one lit it from behind, and every edge moves by a different amount.
A comparator screen shows a circle; parts larger than the circle get moved. A camera system has the same two limits under different names. IMAGE 3 publishes Ø100 mm as the largest single-shot field on IMAGE 3, 25 × 25 mm as the maximum at high magnification on MAX2, and 300 × 200 × 75 mm as the envelope depending on model, with high-precision stage travel of 118 × 118 mm or 218 × 118 mm. MetX is selectable from 300 × 200 × 200 mm to 600 × 800 × 200 mm, with multi-region capture and stitching.
This is the parameter a comparator user feels most, and the one with the least published data anywhere. On a comparator a human aligns a line to a projected edge: repeated readings by one operator cluster tightly, readings between operators do not, and the spread grows on low-contrast edges, rough edges, and late in a shift.
On a vision system the decision is made by an edge extraction algorithm against a grey-level threshold. It is repeatable in a way a human is not, and blind in a way a human is not: it will hold a burr as part of the edge unless told otherwise. IMAGE 3 documentation states that burrs, dirt spots and local defects can be excluded from the fitting range. Ask for repeatability on your edge, not on a gauge block; IMAGE 3 publishes ±0.25 µm, model dependent, as high-precision repeatability on MAX and MAX2.
Once a sensor reads the edge, the floor is set by one division:
pixel equivalent (µm/pixel) = field width in µm ÷ pixel count across that width
A comparator user never needed this number, because the screen and chart defined resolution. IMAGE 3 lists a 1-inch 20-megapixel monochrome CMOS, two on some models, and fields of 6.4 × 4.8 mm / 18.8 × 14.1 mm and 6 × 6 mm / 25 × 25 mm, but the horizontal pixel count is not published. MICRO IMAGE3 gives the other half — 4024 × 3036 — with no single field figure, because the field depends on the objective fitted. Sub-pixel extraction stabilises an edge already resolved; it does not recover one the lighting never made.
A comparator gives a profile. It does not give height, step, thickness, flatness or a 3D contour, and a replacement that only reproduces the profile leaves those unmeasured — fine if nobody relied on them, a serious gap if someone did. IMAGE 3 can be fitted with an 8055-point spectral confocal: 55 mm working distance, ±3 mm range, ±2 µm repeatability, 17 µm spot, 0.02 %F.S. linearity, optional on MAX2. MetX accepts white-light confocal, 3D line laser and probe modules and measures line and surface profile as first-class items. WLI 1000 covers the surface end with vertical resolution below 0.2 nm over a 0–10 mm Z scan.
| Model | Magnification | Contour channel | Field / envelope | Camera | Published figure | Height |
|---|---|---|---|---|---|---|
| **IMAGE 3** | — (sold on field size) | contour backlight + surface + coaxial | Ø100 mm; 25 × 25 mm max at high mag (MAX2); envelope 300 × 200 × 75 mm | 1" 20 MP mono CMOS, ×2 on some models | ±0.25 µm, model dependent; repeatability on MAX / MAX2 | 8055-pt spectral confocal: 55 mm WD, ±3 mm, ±2 µm, 17 µm spot, 0.02 %F.S. (MAX2 optional) |
| **MetX** | PMS continuous-zoom lens | parallel telecentric cold source; 5-ring 8-zone for surface | 300 × 200 × 200 to 600 × 800 × 200 mm; stitching | 1.3 / 2 / 5 MP | — | confocal, line laser, probe; optional |
| **MICRO IMAGE3** | objective 1X–150X; display 1X–8000X | — | 100 × 100 mm stage; 200 × 200 / 300 × 200 / 300 × 300 optional | 4K, 4024 × 3036 | stage repeatability 1 µm | Z travel 30 mm; Z resolution 0.1 µm; 75 mm sample height |
| **LM-1030** | — | GaN green LED, telecentric | 0.3–30 mm; 160 ± 40 mm | HL-CCD | ±2 µm | — |
| **WLI 1000** | per field and lateral resolution | brightfield; PSI / VSI / EDF | 50 × 50 mm XY stage | 1920 × 1200 | vertical resolution < 0.2 nm | Z scan 0–10 mm; ±12° tilt |
The figures were not measured under the same conditions or for the same purpose. The ±0.25 µm on IMAGE 3 is repeatability on a static part; the ±2 µm on LM-1030 is measuring accuracy on a part moving through a beam at 2,400 samples per second. The dashes are informative: IMAGE 3 is sold on field size, not magnification.
IMAGE 3 Series is the direct answer. The part is placed in any orientation, the system recognises position and direction, one press runs the program, and under the right model and recipe the series measures up to 300 locations and 100 workpieces in one program. Surface, coaxial and contour backlight channels let the comparator's silhouette be reproduced on contour backlight while surface features are read separately.
A comparator handles these awkwardly, because a cylinder's silhouette is only informative at the orientation you chose. LM-1030 measures diameter directly: 0.3–30 mm non-contact at ±2 µm, 2,400 samples per second, with a GaN green LED, telecentric optics and HL-CCD continuous exposure across 160 ± 40 mm, and with no high-speed motor or polygon there is nothing rotating to wear. Its published environment is 0 to +50 °C, 35 % to 85 % RH non-condensing, IP64.
MetX Series fits when the drawing now asks for line profile, surface profile, thickness, step or flatness: a selectable range from 300 × 200 × 200 mm to 600 × 800 × 200 mm, grating scales at 0.1 µm or 0.5 µm, and linear, squareness, multi-layer spatial plane and combined compensation across the travel. Both light sources are fitted — the parallel telecentric cold source gives the contour edge, the five-ring eight-zone source lights the surface.
MICRO IMAGE3 covers the case where the question is the condition of an edge rather than a dimension. Objectives 1X–150X on 4024 × 3036 give the close-up, the 100 × 100 mm stage at 1 µm repeatability moves between regions, and 30 mm of Z autofocus at 0.1 µm resolution holds a stepped part in focus.
Do not start from the old datasheet — a comparator's specification of magnification, screen diameter and chart accuracy does not transfer, because the new instrument has no screen and no chart. Write six lines: profile features and tolerances, one line each; which edge you trust and how it was lit; largest envelope and whether stitching is permitted; edge condition rules, since IMAGE 3 excludes burrs and dirt from the fitting range; height items, stated explicitly; and environment and output.
This is a parameter reference, not a compatibility statement. No claim is made that any POMEAS instrument is a drop-in replacement for a specific third-party optical comparator, and none should be inferred. Whether a substitute reproduces the numbers a line accepts depends on parts, fixturing, lighting and recipe, none of which a datasheet settles.
Third-party brand names appear only as recorded in customer-supplied intelligence: manufacturers describing their own installed equipment in public material. No third-party specification is quoted, because none was verified. The count of 7 is evidence about installed base.
Optical resolution is not measurement accuracy. Resolution follows from optics, sensor pitch and wavelength; accuracy additionally depends on edge contrast, illumination stability, positioning, thermal state and algorithm — which matters most here, because the comparator's own accuracy was always partly a property of the operator.
Not identically. A comparator projects a silhouette for a human to read; a vision system extracts an edge from pixels against a threshold.
On MICRO IMAGE3, 1X–8000X is the digital display range and 1X–150X is the objective range. Only the objective figure is optical.
Because diffuse backlight wraps around an edge. A collimated contour source — the parallel telecentric cold source on MetX — reduces that sensitivity.
Repeatability on your edge, not on a gauge block. IMAGE 3 publishes ±0.25 µm, model dependent, as high-precision repeatability on MAX and MAX2.
Yes, by stitching, if the platform supports it. MetX stitches multiple regions with 0.1 µm or 0.5 µm grating scales and compensation across the travel.
Height, step, thickness, flatness and 3D profile. IMAGE 3 offers an 8055-point spectral confocal at ±3 mm; MetX accepts confocal, line-laser and probe modules.
---
If you are specifying a replacement and need measured values for your own conditions — your sensor size, your working distance, your feature size — send us those four numbers and we will run them against the current catalogue and tell you what actually fits, including when nothing does.